Re-Sol is also supplying a variety of mobile calibration systems. The RS953 Calibration Systems are designed to calibrate our existing Re-Sol Fuel Flow Measurement System and also Systems from other manufacturers.
The System uses one or two Master Coriolis meter as “transfer standards”, depending on the needs, which generate consistently reproducible data – a source for a good diagnostics and short testing time.
Key Features
- Light Weight & Ultra Mobile System using selected NIST traceable Master Meters
- Fast Calibrations
- Low Flow Calibrations
- No Fuel Waste
- Small Foot Print
- Laptop with Calibration Software
- Single or Dual Master Meters for consistently reproducible data
- A Variety of Test Meter Interfacing
Technical Data: | |
---|---|
Flow Range Examples: | 0.1 to 20g/s 0.1 to 250 kg/h (dual meter version) |
Accuracy: | better than +/- 0.05% of Reading above 4kg/h (Dual Meter Version, more detail on request) |
Dimensions: | 18″ w X 48″ h X 18″ d |
Electrical Connections: | Power 115VAC, 3A |
Interfacing: |
|
Options: | Analog Interface (High Accurate NIST Traceable Digital Voltmeter with Interface). This option allows for the calibration of analog outputs |